Nextremity Solutions: Medical Device Recall in 2022 - (Recall #: Z-0912-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

Nextra Hammertoe Correction System 3.5 Middle 3.2 Proximal Kit SKU: NX-3532K Orthopedic surgical instrument

Product Classification:

Class II

Date Initiated: February 10, 2022
Date Posted: April 20, 2022
Recall Number: Z-0912-2022
Event ID: 89764
Reason for Recall:

Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery

Status: Terminated
Product Quantity: 285 units
Code Information:

Lot Number: 168125017A UDI: 00817701020042

Distribution Pattern:

Distribution US Nationwide, Japan, and the Netherlands

Voluntary or Mandated:

Voluntary: Firm initiated