Nico Corp.: Medical Device Recall in 2015 - (Recall #: Z-2767-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

NICO Myriad Console Model number NN-7000 The Myriad System is a surgical system specifically designed for morcellation of soft tissue. It primarily consists of a reusable console with foot pedal and a disposable handpiece. For the morcellation and removal of tissue during pelviscopic, laparoscopic, percutaneous and open surgical procedures whenever access to the surgical site is limited.

Product Classification:

Class II

Date Initiated: July 29, 2015
Date Posted: September 23, 2015
Recall Number: Z-2767-2015
Event ID: 72015
Reason for Recall:

Two cables within the console were inadvertently twisted during manufacturing. The individual wires within the cables may break and/or contact one another where the cables terminate into two connectors. When these wires break or contact one another (i.e., short), various types of malfunction may occur with the console. One type of malfunction may result in unexpected aspiration increase.

Status: Terminated
Product Quantity: 6 units
Code Information:

Model number NN-7000 SN: 00231, 00232, 00242, 00245, 00248, 00250

Distribution Pattern:

US Distribution to states of: MD, IN, CA, and NC.

Voluntary or Mandated:

Voluntary: Firm initiated