Nihon Kohden America Inc: Medical Device Recall in 2014 - (Recall #: Z-1979-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs monitoring for multiple patients.

Product Classification:

Class II

Date Initiated: June 3, 2014
Date Posted: July 9, 2014
Recall Number: Z-1979-2014
Event ID: 68459
Reason for Recall:

Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate.

Status: Terminated
Product Quantity: 430
Code Information:

Model Number: CNS-6201 Software Version 02-26 only. Serial Numbers: 00485, 00488, 00489, 00490, 00491, 00361, 00429, 00430, 00432, 00436, 00443, 00447, 00478, 00486, 00546, 00564, 00573, 00574, 00575, 00576, 00592, 00626, 00638, 00643, 00651, 00674, 00675, 00677, 00678, 00679, 00687.

Distribution Pattern:

Worldwide Distribution - US in the state of Oklahoma and in the countries of Bolivia, Chile, Germany, Japan, Korea, Taiwan, Yemen, and South Africa.

Voluntary or Mandated:

Voluntary: Firm initiated