Nobel Biocare Usa Llc: Medical Device Recall in 2020 - (Recall #: Z-1302-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Guided Start Drills/Counterbore NobRpl NP - Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants

Product Classification:

Class II

Date Initiated: March 14, 2019
Date Posted: February 26, 2020
Recall Number: Z-1302-2020
Event ID: 84450
Reason for Recall:

Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.

Status: Terminated
Product Quantity: None
Code Information:

(01)07332747005295(10)73425 Catalog# 32737

Distribution Pattern:

Worldwide distribution - US Nationwide distribution and countries of CA, AU, JP, BR, ZA, RU, TW, HK, IN, CN, SA, NZ, MO, CO, LB, AT, BE, CH, CY, DE, DK, EE, ES, FI, FR, GB, GG, GR, HR, HU, IE, LT, LU, NL, PL, PT, RS, SE.

Voluntary or Mandated:

Voluntary: Firm initiated