Nobel Biocare Usa Llc: Medical Device Recall in 2020 - (Recall #: Z-1316-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Implant Retrieval Instrument Hex & Tri-Ch NP/RP 22mm - Product Usage: They can be used to remove implants, if the implant connection has been damaged. It can be used to remove implants with internal conical connection, internal tri-channel connection or external hex connection.
Class II
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
(01)07332747083767(10)73491 (01)07332747083767(10)73600 (01)07332747083767(10)73296 (01)07332747083767(10)73315 (01)07332747083767(10)73595 Catalog# 37471
Worldwide distribution - US Nationwide distribution and countries of CA, AU, JP, BR, ZA, RU, TW, HK, IN, CN, SA, NZ, MO, CO, LB, AT, BE, CH, CY, DE, DK, EE, ES, FI, FR, GB, GG, GR, HR, HU, IE, LT, LU, NL, PL, PT, RS, SE.
Voluntary: Firm initiated