Northeast Scientific Inc.: Medical Device Recall in 2025 - (Recall #: Z-0018-2026)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2025.
Data Source: FDA.
Product Description:
NES Reprocessed 0.9mm OTW Turbo Elite Laser Atherectomy Catheter. Model Number: R-410-152. The NES Reprocessed Turbo-Elite Laser Atherectomy Catheter is indicated for use in the treatment, including atherectomy, of infrainguinal stenoses and occlusions.
Product Classification:
Class II
Date Initiated: August 29, 2025
Date Posted: October 8, 2025
Recall Number: Z-0018-2026
Event ID: 97524
Reason for Recall:
Potential for breaches in the sterile barrier packaging, compromising sterility assurance.
Status: Ongoing
Product Quantity: 561 units
Code Information:
Model No R-410-152; UDI-DI 00850044399055; All Lot Numbers that are within their labeled shelf life with expiration dates prior to 29AUG2026
Distribution Pattern:
US Nationwide distribution to CA, FL, IL, KS, LA, MD, MI, MO, NC, NJ, NM, NV, NY, PA, TN, TX, VA.
Voluntary or Mandated:
Voluntary: Firm initiated