Northern Digital Inc.: Medical Device Recall in 2015 - (Recall #: Z-2084-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

NDI Disposable Reflective Marker Spheres for Brainlab IGS (Image Guided Surgical) system. Product Usage: Disposable reflective marker spheres are attached to reference arrays and instruments, thus enabling the infrared tracking systems to detect the position of the patient and instruments in the surgical field.

Product Classification:

Class II

Date Initiated: June 10, 2015
Date Posted: July 22, 2015
Recall Number: Z-2084-2015
Event ID: 69804
Reason for Recall:

Inadequate curing of adhesive, may cause the two halves of the spheres to separate.

Status: Terminated
Product Quantity: 524, 700 spheres (41773 = 440,820 spheres; 41774 = 77,760 spheres and 41813 = 6,120 spheres)
Code Information:

1) Catalog Number: 41773 (90 pack) All Lots beginning with the prefix C / All expiration dates; 2) Catalog Number: 41774 (270 pack) All Lots beginning with the prefix C / All expiration dates; and 3) Catalog Number: 41813 (60 pack) All Lots beginning with the prefix C / All expiration dates

Distribution Pattern:

US Nationwide Distribution

Voluntary or Mandated:

Voluntary: Firm initiated