Nova Biomedical Corporation: Medical Device Recall in 2014 - (Recall #: Z-0650-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

Nova StatStrip/StatSensor Hospital Blood Glucose Meter; Nova StatStrip Lactate Hospital Meter System; Nova StatSensor Creatinine Hospital Meter, Rechargeable Lithium Batteries Single: 46626; 4 Pack: 46827; 5 Pack: 50436. For Point-Of-Care usage in the quantitative determination of Glucose in whole blood.

Product Classification:

Class II

Date Initiated: October 10, 2013
Date Posted: January 15, 2014
Recall Number: Z-0650-2014
Event ID: 66398
Reason for Recall:

Revised Battery Instructions for Removal and Expiration Date due to battery swelling and cell leakage

Status: Terminated
Product Quantity: est. 200,000
Code Information:

All serial numbers

Distribution Pattern:

Nationwide Distribution.

Voluntary or Mandated:

Voluntary: Firm initiated