Nova Biomedical Corporation: Medical Device Recall in 2025 - (Recall #: Z-0682-2025)
See the recall detail below. You can also see other recalls from the same firm in 2025.
StatStrip Glucose Ketone Hospital Meter System. Model/Catalog Number: 63683. Prescription use blood glucose ketone meter for near-patient testing
Class I
A software bug within the meter firmware leads to transmission of erroneous glucose and/or ketone patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current test result to the DMS.
Model No 63683. UDI-DI (01)38548063685. Software versions: v0.0.13.10 to v0.0.13.44.
Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.
Voluntary: Firm initiated