NovaSignal Corp.: Medical Device Recall in 2023 - (Recall #: Z-1050-2023)
See the recall detail below. You can also see other recalls from the same firm in 2023.
NovaGuide Intelligent Ultrasound System, Model: NA-RBTSYS1. Comprised of NovaBot Automated Headset (AKA: NeuralBot, Model: NA-RBT1) and Lucid TCD 2.0 Transcranial Doppler Ultrasound (Model: NA-LUCM1)
Class II
Transcranial Doppler/ultrasound system tablet has a faulty hard drive, which will cause the system to freeze and reboot. Initially, the system will reboot to a normal operable state but eventually the system will be rendered inoperable.
System (Tablet)UDI-DI/ Serial Numbers: 00850681007443/ 22040001, 00850681007443/ 22040002, 00850681007443/21021223 (00850681007221/ 21090003), 00850681007443/ 22064227 (00850681007221/ 21090004)
US Nationwide and Worldwide Distribution: AZ, TX, SC, PA, IL. OUS: CA, SE, NL
Voluntary: Firm initiated