NovaSignal Corp.: Medical Device Recall in 2024 - (Recall #: Z-2217-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

NovaGuide 2 Traditional Ultrasound , REF: NSC-TCDNG2

Product Classification:

Class II

Date Initiated: May 6, 2024
Date Posted: July 3, 2024
Recall Number: Z-2217-2024
Event ID: 94673
Reason for Recall:

Due to inadvertently disconnecting grounding cable of ultrasound system could result in electrical shock, electrical interference and/or equipment malfunction.

Status: Ongoing
Product Quantity: 28 systems
Code Information:

UDI: 00850681007405/Serial Numbers: 21092746 21122802 21122803 21092748 21092749 21122804 21123066 23063826 22074381 23063828 22033706 22064206 22074380 21092747 22113115 23063825 21122805 21082646 23063856 23104174 21123068 22103022 22064204 22064205 21123067 22074378 22074379 23033578

Distribution Pattern:

Worldwide - US Nationwide distribution in the states of MA, TX, NY, MN, AZ, AL, CO, CA, OH, MO, FL, TN, NJ, GA, OK, KY, WA and the country of Canada.

Voluntary or Mandated:

Voluntary: Firm initiated