NRT X-RAY A/S: Medical Device Recall in 2023 - (Recall #: Z-1668-2023)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
Intelli-C, Right side suspended Tabletop, REF: 03400000
Product Classification:
Class II
Date Initiated: April 27, 2023
Date Posted: June 7, 2023
Recall Number: Z-1668-2023
Event ID: 92230
Reason for Recall:
There is a potential that the drive shaft in the detector up/down (IDE) movement of the X-Ray machine may break which causes the detector to move until it is stopped by the movement end-stop.
Status: Ongoing
Product Quantity: 32 units
Code Information:
UDI-DI: 05713464000527/Serial Numbers: 627004, 627007, 639001, 639002,639003, 639004, 639007, 639010, 647001, 647004, 647005, 648001, 651007, 658002, 658003, 658006, 658010, 666008, 675001, 675004, 682006, 682010
Distribution Pattern:
US Nationwide distribution in the states of OH, NC, KY, SC, IA, MA, WA, LA, NV, NY.
Voluntary or Mandated:
Voluntary: Firm initiated