NxStage Medical, Inc.: Medical Device Recall in 2015 - (Recall #: Z-0336-2016)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2015.
Data Source: FDA.
Product Description:
NxStage System One S Cycler - Model no. NX1000-3-A (OUS). For home hemodialysis.
Product Classification:
Class II
Date Initiated: October 29, 2015
Date Posted: December 2, 2015
Recall Number: Z-0336-2016
Event ID: 72554
Reason for Recall:
Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed
Status: Terminated
Product Quantity: 133 units
Code Information:
Software version 4.9 and 4.10
Distribution Pattern:
Worldwide Distribution.
Voluntary or Mandated:
Voluntary: Firm initiated