O-Two Medical Technologies, Inc.: Medical Device Recall in 2019 - (Recall #: Z-0978-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
e700 Automatic Transport Ventilator, Model Number 01EVE700
Product Classification:
Class I
Date Initiated: February 12, 2019
Date Posted: April 10, 2019
Recall Number: Z-0978-2019
Event ID: 82178
Reason for Recall:
There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.
Status: Terminated
Product Quantity: 632
Code Information:
EV70001 to EV70658
Distribution Pattern:
The products were distributed to the following US states: AZ, CA, CO, FL, OH, PA, TX, and VA.
Voluntary or Mandated:
Voluntary: Firm initiated