Olympus America Inc.: Medical Device Recall in 2013 - (Recall #: Z-0432-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

TB-0545PC: THUNDERBEAT 5 mm , 45 em, Pistol Grip TB-0535PC: THUNDERBEAT 5 mm, 35 em, Pi stol Grip TB-0545IC: THUNDERBEAT 5 mm, 45 em, In line Grip TB-0535IC: THUNDERBEAT 5 mm, 35 em, Inline Grip TB-0520IC: THUNDERBEAT 5 mm, 20 em, In line Grip TB-051 OIC: THUNDERBEAT 5 mm, I 0 em, In line Grip TB-545FC: THUNDERBEAT 5 mm, 45 em, Front Actuated Grip TB-0535FC: THUNDERBEAT 5 mm, 35 em, Front Actuated Grip The Olympus Thunderbeat Surgical Tissue Management System ("Thunderbeat") is intended to be used for open, laparoscopic (including single-site surgery), and endoscopic surgery to cut (dissect) or coagulate soft tissue or to ligate (seal and cut) vessels.

Product Classification:

Class II

Date Initiated: September 27, 2013
Date Posted: December 11, 2013
Recall Number: Z-0432-2014
Event ID: 66407
Reason for Recall:

Complaints of damage to various models of Thunderbeat Hand Instruments during surgical procedures, including cracking, breaking, or deformation of the components of the probe tip or jaw.

Status: Terminated
Product Quantity: 8709
Code Information:

All sold units

Distribution Pattern:

Worldwide Distribution - USA (nationwide) including Puerto Rico and Internationally to Canada.

Voluntary or Mandated:

Voluntary: Firm initiated