Olympus Corporation of the Americas: Medical Device Recall in 2022 - (Recall #: Z-1263-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2022.
Data Source: FDA.
Product Description:
Olympus URETERO-RENO FIBERSCOPE Model: URF-P6
Product Classification:
Class II
Date Initiated: April 20, 2022
Date Posted: June 22, 2022
Recall Number: Z-1263-2022
Event ID: 90195
Reason for Recall:
Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.
Status: Ongoing
Product Quantity: 2 units
Code Information:
UDI-DI: 04953170340802 Serial Number: 2836210 2834747I
Distribution Pattern:
CA FL MA MD MI NJ OH PR WI
Voluntary or Mandated:
Voluntary: Firm initiated