Olympus Corporation of the Americas: Medical Device Recall in 2022 - (Recall #: Z-1264-2022)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

Olympus URETERO-RENO FIBERSCOPE Model: URF-P6R

Product Classification:

Class II

Date Initiated: April 20, 2022
Date Posted: June 22, 2022
Recall Number: Z-1264-2022
Event ID: 90195
Reason for Recall:

Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.

Status: Ongoing
Product Quantity: 4 units
Code Information:

UDI-DI: 04953170340833 Serial Number: 2836209F 2835978 2837272 2836299

Distribution Pattern:

CA FL MA MD MI NJ OH PR WI

Voluntary or Mandated:

Voluntary: Firm initiated