Olympus Corporation of the Americas: Medical Device Recall in 2023 - (Recall #: Z-2344-2023)
See the recall detail below. You can also see other recalls from the same firm in 2023.
Olympus Endoscope Reprocessor OER-Elite-For use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and their accessories.
Class II
Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible
UDI-DI: 04953170404047 Serial Numbers: 2201375 and the last 4 digits less than 1375
Nationwide
Voluntary: Firm initiated