Product Description:

Soltive SuperPulsed Laser System TFL FIBER CLEAVER, Model Number TFL-AFC

Product Classification:

Class II

Date Initiated: December 4, 2023
Date Posted: January 24, 2024
Recall Number: Z-0717-2024
Event ID: 93613
Reason for Recall:

The strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a risk of contamination.

Status: Ongoing
Product Quantity: 236 units
Code Information:

Lot Number: All units sold prior to 2023

Distribution Pattern:

US Nationwide. Global Distribution

Voluntary or Mandated: