Omega Medical Imaging, Inc.: Medical Device Recall in 2023 - (Recall #: Z-0160-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

Omega Medical Imaging Elevating Monitor Suspension Systems, Part Number 1000-0085

Product Classification:

Class II

Date Initiated: September 8, 2023
Date Posted: November 1, 2023
Recall Number: Z-0160-2024
Event ID: 93123
Reason for Recall:

An actuator separated from the pivot mechanism on a lower monitor boom system column. The lower boom became detached at the point shown and was restrained by the installed safety cable.

Status: Ongoing
Product Quantity: 31 units
Code Information:

all units manufactured between 10-22-2012 and 07-09-2019

Distribution Pattern:

Alabama, California, Illinois, Indiana, Kentucky, Louisiana, Missouri, Montana, North Carolina, Ohio, Oklahoma

Voluntary or Mandated:

Voluntary: Firm initiated