Ortho Clinical Diagnostics Inc: Medical Device Recall in 2018 - (Recall #: Z-0058-2019)
See the recall detail below. You can also see other recalls from the same firm in 2018.
VITROS Immunodiagnostic Products - Folate Reagent Pack(10758750009237) 1513266
Class II
The recalling firm determined that biased results may occur for specific VITROS Immunodiagnostic Products (MicroWell Assays) at biotin concentrations which are lower than indicated in the current Instructions For Use (IFU). Patients who are taking biotin supplements could potentially have biased sample results for affected assays at biotin concentrations lower than indicated in the current IFU.
2470 2480 2490 2501 2510 2521
Worldwide Distribution - US Nationwide distribution, and countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, and Netherlands.
Voluntary: Firm initiated