Ortho-Clinical Diagnostics: Medical Device Recall in 2012 - (Recall #: Z-2067-2012)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

VITROS 4600 Chemistry System Software For use in the in-vitro quantitative, semi quantitative, and qualitative measurement of variety of analytes of clinical interest.

Product Classification:

Class II

Date Initiated: May 21, 2012
Date Posted: August 1, 2012
Recall Number: Z-2067-2012
Event ID: 61910
Reason for Recall:

software automatically switches the lot when a vancomycin (VANC) or valproic acid (VALP) reagent is depleted or the operator manually changes the lot in use causing lack of verification of proper quality assessment.

Status: Terminated
Product Quantity: 77
Code Information:

Version 2.0 & Below

Distribution Pattern:

Worldwide Distribution - USA (nationwide) and the countries of Australia, New Zealand, Brazil, Canada, Chile, Argentina, Paraquay, Ecuador, Colombia, India, Japan, Mexico, Panama, Singapore, Venezuela, UK, Northern Europe, France, Germany, Italy, Spain and Portugal.

Voluntary or Mandated:

Voluntary: Firm initiated