Ortho-Clinical Diagnostics: Medical Device Recall in 2013 - (Recall #: Z-1106-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

VITROS 5,1 FS Chemistry System, Product Code 6801375 and VITROS 5,1 FS Chemistry System Refurbished, Product Code 6801890. Intended for use in the measurement of a variety of chemistry analytes.

Product Classification:

Class II

Date Initiated: February 7, 2013
Date Posted: April 24, 2013
Recall Number: Z-1106-2013
Event ID: 64577
Reason for Recall:

Ortho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS¿ Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal evaluation in its manufacturing facility, it was discovered that an extension spring had insufficient force. If the extension spring does not provide sufficient force, MicroSlides may not be properly positioned within the MicroSlide incubator which has the potential to cause imprecise results. Extension springs are just one component in the slide alignment assembly. OCD's evaluation determined that the probability of an out of specification extension spring alone adversely affecting results is unlikely.

Status: Terminated
Product Quantity: VITROS 5,1 FS: USA: 670 units, Foreign: 1249 units; VITROS 5,1 FS Refurbished: USA: 52 units, Foreign: 2 units
Code Information:

VITROS 5,1 FS Chemistry System Serial Numbers: 34000119 through 34002273; VITROS 5,1 FS Chemistry System Refurbished Serial Numbers: 34000133, 34001240, 34000193, 34001243, 34000212, 34001248, 34000316, 34001251, 34000318, 34001274, 34000337, 34001297, 34000360, 34001311, 34000420, 34001317, 34000451, 34001345, 34000478, 34001378, 34000495, 34001389, 34000497, 34001390, 34000529, 34001396, 34000538, 34001422, 34000552, 34001472, 34000557, 34001504, 34000596, 34001569, 34000601, 34001572, 34000631, 34001589, 34000642, 34001610, 34000799, 34001633, 34000805, 34001668, 34000833, 34001809, 34000851, 34001840, 34000893, 34001850, 34000947, 34001897, 34001172, 34001218

Distribution Pattern:

Worldwide Distribution- USA (nationwide) including Puerto Rico and countries of: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Republic of Panama, Singapore, Spain, and Venezuela.

Voluntary or Mandated:

Voluntary: Firm initiated