Ortho-Clinical Diagnostics: Medical Device Recall in 2013 - (Recall #: Z-1647-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

1) VITROS 5,1 FS Chemistry System (product code 6801375) and 2) VITROS 5,1 FS Chemistry System Refurbished (product code 6801890) Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.

Product Classification:

Class II

Date Initiated: May 6, 2013
Date Posted: July 10, 2013
Recall Number: Z-1647-2013
Event ID: 65207
Reason for Recall:

There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.

Status: Terminated
Product Quantity: Total 101 units (84 units for # 6801375- 26 units domestic & 58 units foreign) and (17 units for # 6801890-15 units domestic, 2 units foreign)
Code Information:

Serial numbers: 1) J34000240, J34000714, J34001562 and J34002163 through J34002245, 2) 34000193, 34000420, 34000478, 34000495, 34000497, 34000529, 34000596, 34000799, 34000805, 34000947, 34001243, 34001422, 34001504, 34001569, 34001589, 34001610, 34001633

Distribution Pattern:

Worldwide Distribution - USA Nationwide including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI & WV, and countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, PR, Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain

Voluntary or Mandated:

Voluntary: Firm initiated