Ortho-Clinical Diagnostics: Medical Device Recall in 2014 - (Recall #: Z-0077-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2014.

Data Source: FDA.
Product Description:

VITROS 4600 Chemistry System (VITROS 5,1 FS System family member), System Product Code 6802445. The VITROS 4600 Chemistry System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.

Product Classification:

Class II

Date Initiated: September 8, 2014
Date Posted: October 22, 2014
Recall Number: Z-0077-2015
Event ID: 69234
Reason for Recall:

Ortho Clinical Diagnostics (OCD) identified an anomaly with Software Version 3.0 and below on the VITROS 4600 Chemistry Systems and VITROS Integrated Systems. Internal testing confirmed that when using calibrator barcode labels supplied with VITROS Chemistry Products Calibrator Kits 1, 2, 3, 4, 6, or 9, an unexpected assay calibration may occur if assay targets are unassigned (i.e., hidden).

Status: Terminated
Product Quantity: USA: 97; Foreign: 232
Code Information:

Software Version 3.0 and below; Serial Numbers J46000012 - J46000456 (serial numbers are sequential; J Numbers are analogous to serial numbers)

Distribution Pattern:

Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.

Voluntary or Mandated:

Voluntary: Firm initiated