Ortho-Clinical Diagnostics: Medical Device Recall in 2014 - (Recall #: Z-1519-2014)
See the recall detail below. You can also see other recalls from the same firm in 2014.
VITROS Chemistry Products Specialty Diluent, REF 855 9825, 6 x 3 mL, IVD --- Ortho-Clinical Diagnostics, Inc. Rochester, NY
Class II
Ortho Clinical Diagnostics (OCD) confirmed the potential for lower than expected C-Reactive Protein (CRP) results only to be generated when using VITROS Chemistry Products Specialty Diluent (Lot F3168) or VITROS Chemistry Products FS Diluent Pack 3 (Lot 01-3266). The VITROS Instructions for Use states that if the if a C-reactive protein concentration exceeds the systems measuring (reportable or dynamic) range, dilute the sample with VITROS Specialty Diluent or a patient sample containing a low concentration of CRP. An initial threefold dilution is recommended. Internal testing determined that results using VITROS CRP Slides have the potential for approximately 51% bias using a 3x dilution factor.
Lot F3168 (exp. 30 Sep 2014)
Nationwide, Puerto Rico, and foreign distribution to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
Voluntary: Firm initiated