Ortho-Clinical Diagnostics: Medical Device Recall in 2014 - (Recall #: Z-1734-2014)
See the recall detail below. You can also see other recalls from the same firm in 2014.
VITROS 4600 Chemistry System, Product Code/Catalog Number 6802445 for in vitro diagnostic use.
Class II
Potential for biased results when a Multiple Window (MW) code is associated with results generated by the VITROS 4600 Chemistry System, the VITROS 5600 Integrated System and the VITROS 5,1 Chemistry System. MW codes may be associated with results from any of the following VITROS Chemistry Products using multiple point (multipoint) rate measurements. The MicroSlide assays that could be affected
Software Version 3.0 and lower; Serial Numbers J46000108 through J46000403
Worldwide distribution including US Nationwide, Puerto Rico, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
Voluntary: Firm initiated