Ortho-Clinical Diagnostics: Medical Device Recall in 2015 - (Recall #: Z-0231-2016)
See the recall detail below. You can also see other recalls from the same firm in 2015.
VITROS 5,1 FS Chemistry System, Catalog Number/Product Code 6801375, Unique Device Identifier No. 10758750001132, and VITROS 5,1 FS Chemistry System Refurbished, Catalog Number/Product Code 6801890, Unique Device Identifier No. 10758750001644. Intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest.
Class II
Software Anomaly during ADD Installation on VITROS 5,1 FS Chemistry Systems using Software Versions 2.2.1 through 2.8. The user modified parameters would not be saved as expected, but remained changed to the default values obtained from the ADD disk.
Software Versions 2.2.1 through 2.8; Catalog # 6801375 Serial Numbers: 34000119 through 34002323; Catalog # 6801890 Serial Numbers: J34000133, J34000193, J34000197, J34000212, J34000285, J34000316, J34000318, J34000337, J34000360, J34000420, J34000451, J34000478, J34000495, J34000497, J34000529, J34000538, J34000552, J34000557, J34000590, J34000596, J34000601, J34000631, J34000642, J34000761, J34000799, J34000805, J34000833, J34000851, J34000893, J34000946, J34000947, J34001006, J34001172, J34001218, J34001240, J34001243, J34001248, J34001251, J34001274, J34001297, J34001311, J34001317, J34001345, J34001378, J34001379, J34001389, J34001390, J34001396, J34001422, J34001472, J34001504, J34001569, J34001572, J34001612, J34001626, J34001633, J34001668, J34001755, J34001809, J34001840, J34001850, J34001875, J34001897.
Worldwide Distribution: US (nationwide) including Puerto Rico, and countries of: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain, and Venezuela.
Voluntary: Firm initiated