Ortho-Clinical Diagnostics: Medical Device Recall in 2016 - (Recall #: Z-0641-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

VITROS 5600 Chemistry System, Catalog Number 6802413, Unique Device Identifier Number 10758750009916, Software Product Code 6802864, Software Version 3.2 and below; IVD. Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.

Product Classification:

Class II

Date Initiated: November 3, 2015
Date Posted: January 20, 2016
Recall Number: Z-0641-2016
Event ID: 72597
Reason for Recall:

Software anomaly may allow VITROS Systems to sample and process assays while the fluids and reagent temperatures are not in the required temperature range for optimal processing. When this intermittent issue occurs, the VITROS Systems continue to operate and process results without notification to the User, possibly producing erroneous results.

Status: Terminated
Product Quantity: Domestic: 1083, Foreign: 999
Code Information:

Software Version 3.2 and below; Serial Numbers J56000024 - J56002309

Distribution Pattern:

Worldwide Distribution - US Nationwide and the countries of Puerto Rico, and foreign distribution to: Argentina, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Israel, Italy, Japan, Kenya, Mexico, Panama, Philippines, Poland, Romania, Russia, Singapore, Spain, Taiwan, Turkey, and Venezuela.

Voluntary or Mandated:

Voluntary: Firm initiated