Ortho-Clinical Diagnostics: Medical Device Recall in 2016 - (Recall #: Z-2804-2016)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

The VITROS¿ Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITROS¿ Chemistry Products HbA1c Reagent Kit, Generation 6 (GEN 6) and Assay Data Disk (ADD), Data Release Versions (DRV) 5873, 5874, 5875, or 5876, is used to calibrate the VITROS 5,1 FS Chemistry System, VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the determination of percent glycated hemoglobin (HbA1c) in human whole blood.

Product Classification:

Class III

Date Initiated: April 19, 2016
Date Posted: September 21, 2016
Recall Number: Z-2804-2016
Event ID: 74009
Reason for Recall:

There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a successful calibration of the assay.

Status: Terminated
Product Quantity: 520 Units
Code Information:

Lot 3155 Product Codes: 6842906 & 6801876

Distribution Pattern:

Worldwide Distribution - US Nationwide in the states of: AL, AR, AZ, CA, CT, DC, FL, GA, IL, IN, KY, LA, MD, MI, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, WA, WV & Puerto Rico;** Foreign: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy & Spain.

Voluntary or Mandated:

Voluntary: Firm initiated