Ortho-Clinical Diagnostics: Medical Device Recall in 2017 - (Recall #: Z-0254-2018)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2017.
Data Source: FDA.
Product Description:
enGen Track System with TCAutomation Software Version 4.2
Product Classification:
Class II
Date Initiated: April 26, 2017
Date Posted: December 20, 2017
Recall Number: Z-0254-2018
Event ID: 77172
Reason for Recall:
A software anomaly with TCA Software V4.2 may potentially cause a delay in reporting of results due to patient samples being routed to an unintended location when samples are reintroduced back onto the enGEN track
Status: Terminated
Product Quantity: 16
Code Information:
Serial numbers: 5GV6W2J, 3RHC23J, 7KSC15J, 953252, CXL941J & JG7BFB5J Product Code 952019-EG, 952020-EG, 952055
Distribution Pattern:
Domestically - GA, IL, MO, NC & NY; Internationally - Canada, England, France, Italy & Spain
Voluntary or Mandated:
Voluntary: Firm initiated