Ortho-Clinical Diagnostics: Medical Device Recall in 2018 - (Recall #: Z-3043-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

VITROS Chemistry Products Cl- Slides, Product Code 844 5207 Product Usage: For in vitro diagnostic use only. VITROS Chemistry Products Cl- Slides quantitatively measure chloride (Cl-) concentration in serum and plasma using VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Product Classification:

Class III

Date Initiated: July 11, 2018
Date Posted: September 19, 2018
Recall Number: Z-3043-2018
Event ID: 80697
Reason for Recall:

The slides do not meet the maximum allowable interference (MAI) claims for triglyceride interference when using serum samples. Triglyceride interference on the Cl- Slide was observed around a triglyceride concentration 600mg/dL, however the current triglyceride interference claim states interference should not occur below 800mg/dL of triglyceride.

Status: Terminated
Product Quantity: 128196
Code Information:

GEN 5 and below

Distribution Pattern:

Worldwide Distribution - US Nationwide and in the countries of: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, and United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated