Product Description:

Rival Reduce Kit Plates, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Part Number: 000-K-012C

Product Classification:

Class II

Date Initiated: April 6, 2018
Date Posted: June 13, 2018
Recall Number: Z-2118-2018
Event ID: 79819
Reason for Recall:

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Status: Terminated
Product Quantity: 50
Code Information:

UDI: 18056099648608, Lot: B1107421,

Distribution Pattern:

U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,

Voluntary or Mandated:

Voluntary: Firm initiated