Orthofix, Inc: Medical Device Recall in 2018 - (Recall #: Z-2131-2018)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2018.
Data Source: FDA.
Product Description:
MTP Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (002-B-01001); b) Medium (002-B-02002); c) Large (002-B-03003)
Product Classification:
Class II
Date Initiated: April 6, 2018
Date Posted: June 13, 2018
Recall Number: Z-2131-2018
Event ID: 79819
Reason for Recall:
Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Status: Terminated
Product Quantity: N/A
Code Information:
(UDI)/Lot: a) (18056099647557) B1098799, B1115575; (18056099647564) B1113628; (18056099647571) B1098800
Distribution Pattern:
U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
Voluntary or Mandated:
Voluntary: Firm initiated