Orthofix, Inc: Medical Device Recall in 2018 - (Recall #: Z-2133-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

TN Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a) Small (002-F-01020); b) Medium (002-F-01025); c) Large (002-F-01030)

Product Classification:

Class II

Date Initiated: April 6, 2018
Date Posted: June 13, 2018
Recall Number: Z-2133-2018
Event ID: 79819
Reason for Recall:

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

Status: Terminated
Product Quantity: 24
Code Information:

(UDI)/Lot: a) (18056099647663) B1099060; b) (18056099647670) B1099062; c) (18056099647687) B1099066

Distribution Pattern:

U.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,

Voluntary or Mandated:

Voluntary: Firm initiated