Orthofix Srl: Medical Device Recall in 2019 - (Recall #: Z-0385-2020)
See the recall detail below. You can also see other recalls from the same firm in 2019.
ORTHOFIX Catalogue Number: ref 1-1355001, DRILL BIT D.2,7 MM L.127 MM TIN COATED- QUICK CONNECT, RX ONLY, Non Sterile, UDI: (01)18054242S10918
Class II
As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Lot # 300244, 300243, 289816, 273348
US: AZ, CA,FL,GA,HI,IA,IL,IN,KY,LA,MD,MI,MO,MS,NC,NY,OH,OK,OR,PA,SC,TN,TX,UT,VA,WA, WI, WV, Puerto Rico OUS: Kuwait, Australia, Saudi Arabia, United Arab Emirates, Ireland, Japan, Spain, Poland, Switzerland, Singapore, Peru, Brazil, Germany, Great Britain, France, Italy, Australia, South Africa, Belgium, Chile, The Netherlands, Lithuania, Sri Lanka, Denmark, Israel
Voluntary: Firm initiated