Orthovita, Inc., dBA Stryker Orthobiologics.: Medical Device Recall in 2013 - (Recall #: Z-1389-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Stryker Orthobiologics Aliquot Delivery System Manual Surgical Instrument - Aliquot Plunger Delivery System, 4", Part Number 2090-0501 Product Usage: Facilitate placement of bone cement in weakened or diminished bone

Product Classification:

Class II

Date Initiated: November 10, 2012
Date Posted: June 5, 2013
Recall Number: Z-1389-2013
Event ID: 63204
Reason for Recall:

There have been a few reported cases of the catheter breaking when removing the Aliquot catheter from the needle during vertebroplasty procedures. This has been attributed to bending of the catheter during the retraction from the needle. It was determined to be use errors. The correction is limited to a revision of the Directions for Use.

Status: Terminated
Product Quantity: 411
Code Information:

Manufacturer Part Number 2090-0501, Lot Numbers A1109035 and Q1006002

Distribution Pattern:

USA Nationwide Distribution in the states of AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MD, ME, MI, MO, NC, ND, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, and WI.

Voluntary or Mandated:

Voluntary: Firm initiated