Oscor, Inc.: Medical Device Recall in 2016 - (Recall #: Z-0794-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2016.

Data Source: FDA.
Product Description:

OSCOR ADELANTE BREEZEWAY 8F C55¿ S79CM D84CM Catalog # AB081044 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.

Product Classification:

Class II

Date Initiated: October 5, 2016
Date Posted: December 28, 2016
Recall Number: Z-0794-2017
Event ID: 75785
Reason for Recall:

Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use.

Status: Terminated
Product Quantity: 991
Code Information:

Lot #'s: C1-12152 Exp. November 30, 2018, C1-09518 Exp. December 21, 2017, C1-09748 Exp. February 16, 2018, C1-09899 Exp. March 12, 2018, C1-10040 Exp. April 5, 2018, C1-10040 Exp. April 5, 2018, C1-10149 Exp. April 23, 2018, C1-10330 Exp. May 11, 2018, C1-10329 Exp. May 11, 2018, C1-10382 Exp. May 17, 2018, C1-11571 Exp. September 25, 2018, C1-11571 Exp. September 25, 2018, and OR-04757 Exp. December 22, 2018

Distribution Pattern:

Worldwide Distribution - US including CA, NJ, MD, TX, FL, NC, MO, and Puerto Rico. Internationally to AUSTRIA, BOLIVIA, CANADA, CHILE, CYPRUS, NETHERLANS, EGYPT, GERMANY, HONG KONG, ISRAEL, ITALY, SWITZERLAND, and VIETNAM

Voluntary or Mandated:

Voluntary: Firm initiated