Ossur H / F: Medical Device Recall in 2015 - (Recall #: Z-1288-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

LP Rotate Foot System prosthetic device for fitting lower extremity amputations. -Low profile foot module with adjustable shock absorber and rotation with progressive stiffening -Build height: with male pyramid adapter 145 - 153mm (5 11/16 - 6 ) with female pyramid adapter 153 165mm (6 - 6 1/2) -Male pyramid adapter -Female pyramid adapter -Maintenance free, no lubrication required -Maximum weight: 147kg (325lbs) -Available in sizes 22-30

Product Classification:

Class II

Date Initiated: March 4, 2015
Date Posted: March 25, 2015
Recall Number: Z-1288-2015
Event ID: 70648
Reason for Recall:

The threads of the outer tube of the shock module of LP Rotate are too short, resulting in insecure assembly. Use of the product may cause the patient to fall with continued use, as the shock module can become loose from the foot blade.

Status: Terminated
Product Quantity: 2,194 (1,464 in U.S. and 730 internationally)
Code Information:

LP Rotate, sizes 25-27 cat.1-6. Product numbers LRP062, LRF062, LRPU62, LRFU62. All from manufacturing lot number HF141022

Distribution Pattern:

Nationwide within the U.S. and internationally to the following foreign countries; Belgium, United Kingdom, Germany, Czech Republic, France, Israel, Poland, Spain, Australia, Japan, Mexico, Bulgaria, Canada, South Korea, Sweden, Finland,

Voluntary or Mandated:

Voluntary: Firm initiated