Ossur H / F: Medical Device Recall in 2023 - (Recall #: Z-1906-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

Icelock 125 Ratchet, REF: L-125000; Icelock 600XM Ratchet, REF: L-621200; Icelock 621 Ratchet, REF: L-621000; Icelock 621 Ratchet Adaption Kit, REF: L-621100; Ratchet Lock Body, REF: L-692020

Product Classification:

Class II

Date Initiated: May 12, 2023
Date Posted: June 21, 2023
Recall Number: Z-1906-2023
Event ID: 92207
Reason for Recall:

A mechanical lock used to connect and release a locking liner into a prosthetic socket have an attachment pin and lock that can wear down faster than expected, leading to the potential for the pin to disengage from the lock, leading to loss of suspension of the external prosthetic system that replaces a missing lower limb, which can lead to falling and injury.

Status: Ongoing
Product Quantity: 4,190
Code Information:

REF/UDI-DI: L-125000/05690967566142; L-621200/05690967596590; L-621000/05690967118334; L-621100/05690967155322; L-692020/05690967128548. Lots: HF210205, HF210519, HF211008, HF220303, MX210518, MX210816, MX210823, MX211022, MX220316.

Distribution Pattern:

US: SC, FL,NM, PA, VA, CA, TX, MT, UT, OH, NV, KS, MI, GA, IA, WA, TN, AZ, MN, NC, IN, MS, AL, NY, ID, DE, OK, MA, WI, OR, KY, MD, CO, WV, LA, IL, NJ, CT, MO, NE, WY, AR, SD, PR

Voluntary or Mandated:

Voluntary: Firm initiated