Palodex Group Oy: Medical Device Recall in 2020 - (Recall #: Z-0841-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
KaVo Scan eXam SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
Class II
Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.
Serial Number: TBD (range from KJ1208178 to KJ1314403) PSU's having nine-digit serial numbers beginning with four digits as follows: 1143XXXXX 1221XXXXX 1224XXXXX and 1229XXXXX
US: WI and PA OUS: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CZECH REPUBLIC, DENMARK, ECUADOR, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KOREA, REPUBLIC OF, KUWAIT, MALAYSIA, MALTA, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PAKISTAN, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, REPUBLIC OF CHINA, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM
Voluntary: Firm initiated