Pentax Medical Company: Medical Device Recall in 2016 - (Recall #: Z-2078-2016)
See the recall detail below. You can also see other recalls from the same firm in 2016.
Ureteroscope Intended to provide optical visualization of and therapeutic access to the Urinary Track. This anatomy includes but is not restricted to the organs, tissues and subsystems: Urethra, Bladder and Renal Pelvis. The instrument is introduced via the urethra when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.
Class II
Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel.
Model #'s FUR-9P and FUR-9RBS
Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam
Voluntary: Firm initiated