Pentax of America Inc: Medical Device Recall in 2017 - (Recall #: Z-3068-2017)
See the recall detail below. You can also see other recalls from the same firm in 2017.
Ultrasound Video Gastroscope, Model EG-3870UTK. The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.
Class II
Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.
H110516, H110619, H110635, H110643, H110624, A110407, A110411, H110507, A110352, A110444, A110417, A110419, H110493, H110494, H110522, H110584, H110586, H110590, H110591, H110559, A110471, A110474, A110355, A110359, H110606, H110566, H110568, H110498, H110503, H110509, H110510, H110637, H110560, H110514, A110427, A110372, A110438, H110587, H110598, H110600, A110439, A110440, A110443, A110450, A110448, A110451, A110453, H110542, H110569, A110445, A110454, H110565, A110356, A110357, H110592, H110593, H110595, A110346, H110508, A110461, H110620, H110623, H110641, H110642, H110613, A110460, A110380, A110463, H110577, A110457, A110401, H110610, H110611, H110502, H110511, H110518, H110519, H110578, H110557, H110561, A110422, A110432, H110512, H110626, H110628, H110629, H110633, H110634, H110539, H110544, A110341, A110418, H110596, H110604, A110387, A110388, H110601, H110615, H110571, H110574, H110599, H110506, A110350, A110398, H110575, A110449, A110360, A110363, A110368, H110554, H110515, H110567, A110410, A110414, A110485, H110553, H110563, H110589, A110376, A110384, H110549, H110564, H110556, H110536, H110646, H110533, H110543, A110458, A110459, A110383, H110538, H110499, A110344, A110345, A110348, A110452, A110391, A110475, A110483, A110486, A110488, A110477, A110479, A110480, A110481, A110482, A110375, A110406, A110470, A110429, A110435, H110618, A110393, A110394, A110447, A110462, A110466, H110555, H110631, H110636, H110639, H110524, H110525, H110526, H110530, H110531, H110532, H110534, H110535, A110476, A110416, A110420, A110379, A110381, A110385, A110484, A110492, H110640, H110579, H110581, H110572, H110573, A110425, A110431, A110441, A110442, H110517, A110426, A110433, A110434, A110353, H110614, H110616, A110464, A110465, H110520, H110523, H110550.
Nationwide
Voluntary: Firm initiated