Philips And Neusoft Medical Systems Co., Ltd.: Medical Device Recall in 2013 - (Recall #: Z-1166-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

NeuViz Dual series CT Scanner System, Part number (PN): 989605651321. This is a computed X-ray, Tomography system. NeuViz Dual Multi-slice CT Scanner System is intended to produce cross-section images of head and body by computer reconstruction of X-ray transmission data taken at different angles.

Product Classification:

Class II

Date Initiated: March 4, 2013
Date Posted: May 1, 2013
Recall Number: Z-1166-2013
Event ID: 64783
Reason for Recall:

User error of the NeuViz 16 or NeuViz Dual may cause personnel injuries. There was a report where a finger was pinched when the table was moved backward while the tabletop was controlled by another person from the gantry left control panel.

Status: Terminated
Product Quantity: 25 units
Code Information:

Serial numbers: 400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013.

Distribution Pattern:

Nationwide distribution: USA including states of: California, Connecticut, North Carolina, Nebraska, Ohio, South Carolina, Texas, and Puerto Rico.

Voluntary or Mandated:

Voluntary: Firm initiated