Philips Electronics North America Corporation: Medical Device Recall in 2017 - (Recall #: Z-0015-2018)
See the recall detail below. You can also see other recalls from the same firm in 2017.
Philips Healthcare Brilliance 64 System, X-Ray, Tomography, Computed These computed tomography x-ray systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.
Class II
During a bolus tracking procedure, no images were generated when the scan was completed, and the raw data file was not available for offline reconstruction. There is a discrepancy between the calculated reconstruction length and the actual scan length that results in the inability to reconstruct raw data. The operator may choose to rescan the patient.
Software version 4.16
USA (nationwide) Distribution
FDA Mandated