Philips Electronics North America Corporation: Medical Device Recall in 2017 - (Recall #: Z-2528-2017)
See the recall detail below. You can also see other recalls from the same firm in 2017.
M3150 Information Center Local Database; Models: 862249, 865001, 865003, 865004, 865007, 865009, 865011, 865012, 865091, 865092, 865093, 865094, 865095, 865096, 865098, 865099, 865102, 865104, 865105, 865125, 865127. The intended use of the Information Center Software is to display physiologic waves, parameters and trends, formal data for strip recordings and printed reports and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its database. An additional intended use is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors
Class II
Potential for early component failure.
This action affects certain customers who were shipped an affected Philips Intellivue Information Center unit between June 9, 2008 and May 31, 2009. Affected products have a PC serial number range between 2UA8300XYZ and 2UA84510QW. However, only a subset of these products are affected.
Worldwide Distribution - US (Nationwide) OUS: CANADA ARGENTINA AUSTRALIA AUSTRIA BAHRAIN BELGIUM BRAZIL CHILE CHINA COLOMBIA CZECH REPUBLIC FINLAND FRANCE GERMANY GREECE INDIA INDONESIA IRELAND ITALY JAPAN KENYA KOREA, REPUBLIC OF NETHERLANDS PHILIPPINES POLAND SPAIN SWEDEN SWITZERLAND THAILAND TURKEY UNITED KINGDOM
Voluntary: Firm initiated