PHILIPS HEADQUARTERS CAMBRIDGE: Medical Device Recall in 2023 - (Recall #: Z-1623-2023)
See the recall detail below. You can also see other recalls from the same firm in 2023.
Incisive CT- Whole-body computed tomography (CT) X-Ray System Model Number: 728143 (OUS)
Class II
Potential for Incorrect Image Orientation resulting images may be flipped or reversed result in misdiagnosis, incorrect treatment of a condition, or additional radiation exposure if a rescan is required
UDI-DI: (01)00884838085015(21) Serial Numbers: 75592542 81629039 82435668 84375501 85146962 85469672 86188405 86188817 86820074 87367597 92174932 92622328 92637836 92680372 92823844 92870027 92889235 94581069 95050209 95055576 96094114 96462378 Serial Numbers: 75592542 81629039 82435668 84375501 85146962 85469672 86188405 86188817 86820074 87367597 92174932 92622328 92637836 92680372 92823844 92870027 92889235 94581069 95050209 95055576 96094114 96462378
AR, CA, OK, TX Foreign: Country Argentina Australia Austria China Colombia Czech Republic Dominican Rep France Germany Guadeloupe Hong Kong India Japan Kuwait Latvia Lithuania Netherlands Norway Portugal Romania Spain Switzerland Taiwan Thailand United Kingdom Vietnam
Voluntary: Firm initiated