Philips Healthcare Inc.: Medical Device Recall in 2012 - (Recall #: Z-2494-2012)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

Philips Digital Diagnost with patient carriage Model: 712050 with serial number range from 00 000 00 to 09 000 543. Product Usage: Stationary radiographic System

Product Classification:

Class II

Date Initiated: September 12, 2012
Date Posted: October 10, 2012
Recall Number: Z-2494-2012
Event ID: 63165
Reason for Recall:

Hook does not securely hold the footplate in vertical position, causing the footplate to disengage and may fall on patient or operator

Status: Terminated
Product Quantity: 536 units
Code Information:

Serial number range (patient carriage) from: 00 000 00 to 09 000 543.

Distribution Pattern:

Worldwide Distribution - US Nationwide and the countries of :Australia, Austria, Bahrain, Belgium, Canada, Chile, China, Czech Republic, Denmark, Finland, , France, Germany, Ireland, Italy, Japan, Netherlands, New Zealand, Norway, Poland, Qatar, Romania, Russia, Saudi Arabia, Sweden, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan. Thailand, Turkey,United Arab Emirates and United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated