Philips Healthcare Inc.: Medical Device Recall in 2013 - (Recall #: Z-0767-2013)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

Phillips Easy Diagnost Eleva DRF. Model Number: 706034 Product Usage: Universal Fluoroscopic Applications

Product Classification:

Class II

Date Initiated: January 16, 2013
Date Posted: February 13, 2013
Recall Number: Z-0767-2013
Event ID: 64127
Reason for Recall:

The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system.

Status: Terminated
Product Quantity: 14 units
Code Information:

Serial Numbers 362600, 362601, 364209, 364210, 375245, 375850, 376439, 384396, 387492, 398185, 399440, 399621, 399623, 402409

Distribution Pattern:

US Nationwide Distribution including the states of FL, CA, WA, AZ, NC, KY, DE, AZ and MO.

Voluntary or Mandated:

Voluntary: Firm initiated