Philips Healthcare: Medical Device Recall in 2016 - (Recall #: Z-1719-2016)
See the recall detail below. You can also see other recalls from the same firm in 2016.
Philips Healthcare Ingenuity Elite Computed Tomography X-Ray System
Class II
Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm
Software v. 4.13, 4.14 and 4.15.
US Distribution - AZ, PA, FL, DC, W VA, NC, SC, AR, OK, OH, NY, NV, TX, PA, IN, CA, MN, MA, DC, MS, MD, OR, IL, CO, MO, VT, IA, AL, LA, GA and WA.
Voluntary: Firm initiated